FDA-Device2014-12-31Class I

ADULT Radiotransparent Electrode, Part number C100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.

Heart Sync, Inc.
Hazard

Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the

Sold states
Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and South Korea.
Affected count
sold in boxes of ten electrodes: 4663.8 boxes; 46638 units total
Manufactured in
5643 Plymouth Rd, Ann Arbor, MI, United States
Products
ADULT Radiotransparent Electrode, Part number C100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0583-2015