FDA-Device2022-02-16Class II
QC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms
Microbiologics Inc
Hazard
Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Sold states
Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Democratic People's Republic of Korea, Republic of Korea, Kuwait, Lebanon, Malaysia, Netherlands, New Zealand, Panama, Peru, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Affected count
139 units
Manufactured in
200 Cooper Ave N, N/A, Saint Cloud, MN, United States
Products
QC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0588-2022Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief