FDA-Device2014-01-08Class II
Philips HeartStart MRx Monitor/Defibrillator Models: M3536A Options A20 - A27, HeartStart MRx ALS Monitor (Grey) The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Philips Medical Systems, Inc.
Hazard
Error in the printing of the therapy energy setting labels on Philips HeartStart M3536A MRx Monitor/Defibrillators. The first energy setting is labeled 10 when it should be labeled1-10
Sold states
Worldwide Distribution - US Distribution including the states of IN, MS, TX and VA., and the countries of DENMARK, FRANCE, GERMANY, SPAIN and SWITZERLAND.
Affected count
93 units
Manufactured in
3000 Minuteman Rd, Andover, MA, United States
Products
Philips HeartStart MRx Monitor/Defibrillator Models: M3536A Options A20 - A27, HeartStart MRx ALS Monitor (Grey) The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0593-2014Don't want to check this manually?