FDA-Device2016-01-13Class II
Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.
Siemens Medical Solutions USA, Inc
Hazard
Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.
Sold states
Nationwide Distribution including: AL, AZ, CA, CO, FL, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NC, NY, OH, OK, PA, SC, TN, TX, VA, and WI.
Affected count
73 units
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0597-2016Don't want to check this manually?