FDA-Device2018-12-19Class II

Salem Sump Dual Lumen Stomach Tube 10 Fr/Ch (3.3 mm) x 36" (91 cm), Product Code 8888264911 Product Usage: The Salem Sump Tubes are intended for decompression and drainage of stomach contents following injury, illness, or surgery in patients with intestinal obstruction, and to prevent the distention of the stomach.

COVIDIEN LLC
Hazard

This voluntary recall is being conducted due to incorrect packaging. In one lot of Salem Sump Dual Lumen Stomach Tubes the 10Fr product was packaged in a 16Fr pouch. The use of a 10 Fr Salem Sump Tube when a 16Fr is needed could result in decreased fluid collection, with the potential to affect patient care, particularly in an emergency situation where the correct size is not readily available.

Sold states
Worldwide Distribution - US Nationwide and foreign distribution to Canada.
Affected count
1522
Manufactured in
15 Hampshire St, N/A, Mansfield, MA, United States
Products
Salem Sump Dual Lumen Stomach Tube 10 Fr/Ch (3.3 mm) x 36" (91 cm), Product Code 8888264911 Product Usage: The Salem Sump Tubes are intended for decompression and drainage of stomach contents following injury, illness, or surgery in patients with intestinal obstruction, and to prevent the distention of the stomach.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0597-2019