FDA-Device2016-01-13Class II

Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.

Elekta, Inc.
Hazard

When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.

Sold states
Worldwide Distribution - US Nationwide in the states of CA, NY, PA, and the countries of Austria, Belarus, Canada, China, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Netherlands, Norway, Poland, South Korea, Spain, Sweden and United Kingdom.
Affected count
154
Manufactured in
400 Perimeter Center Ter NE Ste 50, N/A, Atlanta, GA, United States
Products
Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0604-2016