FDA-Device2018-12-19Class II

Discovery NM/CT 670 ES, model 5376204-70-57

GE Healthcare, LLC
Hazard

When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.

Sold states
CA, DC, OH, UT, WI, Canada
Affected count
2 units
Manufactured in
3000 N Grandview Blvd, N/A, Waukesha, WI, United States
Products
Discovery NM/CT 670 ES, model 5376204-70-57

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0607-2019