FDA-Device2021-02-03Class II

REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE

Enztec Limited
Hazard

The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.

Sold states
CA
Affected count
312 Femoral Impactors
Manufactured in
26 Dakota Crescent, Sockburn, Christchurch, New Zealand
Products
REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0612-2020

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief