FDA-Device2020-12-23Class II

Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Usage: is a device for measuring the refractive power of contact lenses and spectacle lenses and displaying measurement values for sphere, cylinder and prism.

Carl Zeiss Meditec AG
Hazard

Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

Sold states
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia, Canada, Colombia, Croatia, Cyprus, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Latvia, Luxembourg, Malaysia, Maldives, Netherlands, Portugal, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Turkey, and United Kingdom.
Affected count
603 units
Manufactured in
Carl-Zeiss-Promenade 10, N/A, Jena, N/A, Germany
Products
Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Usage: is a device for measuring the refractive power of contact lenses and spectacle lenses and displaying measurement values for sphere, cylinder and prism.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0613-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief