FDA-Device2024-12-11Class II
EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
Hazard
High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).
Sold states
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.
Affected count
1,770 units
Manufactured in
3500 Corporate Pkwy, Center Valley, PA, United States
Products
EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0620-2025More Olympus Corporation of the Americas recalls
- FDA-Device2026-04-08Olympus SOLTIVE Premium laser system recalled for defective power supply
- FDA-Device2026-04-08Olympus SOLTIVE Pro SuperPulsed Laser System recalled for power supply defect
- FDA-Device2026-04-01Olympus PKS Cutting Forceps recalled for defective welds causing jaw breakage
- FDA-Device2026-04-01Olympus MAJ-1443 and MAJ-1444 endoscope suction valves no longer compatible with OER-Pro and OER-Elite reprocessors