FDA-Device2016-01-20Class II

Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagnostic procedures.

Bayer Healthcare
Hazard

The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail

Sold states
Nationwide Distribution to the states of : AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MI, MN, MO, NC, NE, NH, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT and WI.
Affected count
198,598
Manufactured in
1 Bayer Dr, N/A, Indianola, PA, United States
Products
Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagnostic procedures.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0627-2016