FDA-Device2013-01-16Class III

Immunoradiometric assay. Device is an immunoradiometric assay designed for the quantitative measurement of thyroid stimulating hormone (thyrotropin, TSH) in serum. It is intended strictly for in vitro diagnostic. Used as an aid in the assessment of thyroid status.

Siemens Healthcare Diagnostics
Hazard

The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 (level G) lot 035. The affected calibrator vials can be identified by a darker yellow color than normal. The TSI3 (level A), TSI7 (level E), and TSIX (level H)

Sold states
Worldwide Distribution - USA including CA, GA, IL, IN, MI, MN, NY, and Puerto Rico. Internationally to Agentina, Bolivia, Brazil, Canada, Chile, Guatemala, South Korea, Philippines, Paraguay, Taiwan, and Uraguay,
Affected count
Lot 742= 8 units all foreign consignees) Lot 743=7 units (all foreign consignees)
Manufactured in
511 Benedict Ave, Tarrytown, NY, United States
Products
Immunoradiometric assay. Device is an immunoradiometric assay designed for the quantitative measurement of thyroid stimulating hormone (thyrotropin, TSH) in serum. It is intended strictly for in vitro diagnostic. Used as an aid in the assessment of thyroid status.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0648-2013