FDA-Device2014-01-15Class II
Product Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device with BZO, Reditest On-Site Oral Fluid Drug Screen Device & XALEX One Step Rapid Drug Test. Type of Packaging: 25 individually pouched test devices & collectors in a labeled kit box. The OrALert Oral Fluid Drug Screen Device for AMP/mAMP/COC/OPI/THC/PCP/BZO is an immunoassay based on the principle of competitive binding. Drugs that may be pre
Alere San Diego, Inc.
Hazard
Alere initiated this recall of certain lots of the Oral Fluid Drug Screen Device because the sponge may become dislodged from the Saliva Collector handle prior to or during the oral fluid sample collection process. If the sponge becomes dislodged during the specimen collection process, the donor may inadvertently swallow the sponge which could present a choking risk.
Sold states
US Distribution: including states of: AL, CA, PA, and VA.
Affected count
280,710
Manufactured in
9975 Summers Ridge Rd, San Diego, CA, United States
Products
Product Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device with BZO, Reditest On-Site Oral Fluid Drug Screen Device & XALEX One Step Rapid Drug Test. Type of Packaging: 25 individually pouched test devices & collectors in a labeled kit box. The OrALert Oral Fluid Drug Screen Device for AMP/mAMP/COC/OPI/THC/PCP/BZO is an immunoassay based on the principle of competitive binding. Drugs that may be present in the oral fluid specimen compete against their respective drug conjugates for binding sites on their specific antibody.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0653-2014Don't want to check this manually?