FDA-Device2024-01-17Class II

Oscor Intro Kit, 14Fr x 13 and 25 cm Part of Pump Set (0048-0040)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3025

Abiomed, Inc.
Hazard

Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis

Sold states
Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK
Affected count
1 unit within kit
Manufactured in
24 Cherry Hill Dr, Danvers, MA, United States
Products
Oscor Intro Kit, 14Fr x 13 and 25 cm Part of Pump Set (0048-0040)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3025