FDA-Device2024-01-17Class II
Abiomed KIT, 14 Fr Introducer, 13 & 25 cm, ABMD Part of Pump Set (0048-0024, 0048-0032)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3056
Abiomed, Inc.
Hazard
Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis
Sold states
Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK
Affected count
2 units within kit
Manufactured in
24 Cherry Hill Dr, Danvers, MA, United States
Products
Abiomed KIT, 14 Fr Introducer, 13 & 25 cm, ABMD Part of Pump Set (0048-0024, 0048-0032)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3056
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0654-2024Don't want to check this manually?