FDA-Device2022-03-09Class II
Digital Angiography System Shimadzu Trinias Generator Model: D150GC-40
Shimadzu Medical Systems
Hazard
Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.
Sold states
U.S.: CA, FL, GA, IL, IN, KY, LA, MI, MS, OH, PA, SC, TN, TX and WA O.U.S.: Not provided.
Affected count
28 systems
Manufactured in
20101 S Vermont Ave, Torrance, CA, United States
Products
Digital Angiography System Shimadzu Trinias Generator Model: D150GC-40
Official notice