FDA-Device2021-01-20Class II

Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno

Siemens Medical Solutions USA, Inc
Hazard

When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits.

Sold states
Worldwide-US Nationwide
Affected count
5602
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno