FDA-Device2025-12-03Class IIPROCESSING DEFECT

IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML 126 IN; Catalog Number: 490283.

OTHERNationwide distribution
Hazard

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Sold states
Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Affected count
65,904 units
Manufactured in
824 12th Ave, N/A, Bethlehem, PA, United States
Products
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML 126 IN; Catalog Number: 490283.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0656-2026

More B Braun Medical Inc recalls

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief