FDA-Device2016-01-27Class II
VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequency of seizures.
Cyberonics, Inc
Hazard
Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.
Sold states
Worldwide Distribution -- United States, Austria, Belgium, Croatia, Czech Republic, Finland, France, Germany, Iceland, Italy, Netherlands, Norway, Poland, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Cyprus, Israel, Kuwait, Lebanon, Oman, Qatar, Saudi Arabia, and United Arab Emirates.
Affected count
4,935 units
Manufactured in
100 Cyberonics Blvd, N/A, Houston, TX, United States
Products
VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequency of seizures.
Official notice