FDA-Device2019-12-18Class II
Radiofrequency Grounding Pad, REF RF-DGP-L
Abbott Medical
Hazard
Affected lots were manufactured with the protective release liner in an incorrect orientation. When attempting to remove this disposable release liner, a clear film may be left on the grounding pad which may interfere with patient skin contact and lead to uneven heating.
Sold states
nationwide
Affected count
8940 devices
Manufactured in
5050 Nathan Ln N, N/A, Plymouth, MN, United States
Products
Radiofrequency Grounding Pad, REF RF-DGP-L
Official notice