FDA-Device2024-12-18Class II
MICRO-X Rover Mobile X-ray System, # MXU-RV35
Micro-X Ltd.
Hazard
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
Sold states
Worldwide and US Nationwide Distribution: NJ, PR, CA
Affected count
5
Manufactured in
Unit 14 6 Mab Eastern Promenade, Tonsley, Australia
Products
MICRO-X Rover Mobile X-ray System, # MXU-RV35
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0664-2025Don't want to check this manually?