FDA-Device2016-12-07Class II

Monaco TRP System The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Elekta, Inc.
Hazard

Contours are not handled correctly if the CT dataset contains slices with DICOM Z coordinates of xx.x50mm.

Sold states
Worldwide Distribution - US including AK, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MD, MA, MI, MN, MO, MT, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, SD, TX, UT, VA, WA, WI, Puerto Rico and Internationally to Algeria, Antiqua & Barbuda, Argentina, Australia, Austria, Bahrain, Belarus, Botswana, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Cuba, Cyprus, Czech Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Laos, Lithuania, Malaysia, Malta, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Turkey, Unite Kingdom, Ukraine, Venezuela, Vietnam, and Zimbabwe.
Affected count
2148
Manufactured in
400 Perimeter Center Ter NE Ste 50, Atlanta, GA, United States
Products
Monaco TRP System The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0668-2017

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief