FDA-Device2024-01-17Class II
Boston Scientific Model 3120 ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) installed with Model 2892 Programmer Application Software used for the following pacemaker products: ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus. ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus.
Boston Scientific Corporation
Hazard
The software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will display certain diagnostic dates incorrectly, with a year in the 1990s.
Sold states
Distribution was nationwide, including Guam, Puerto Rico, and the U.S. Virgin Islands. There was also government/military distribution. Foreign distribution was made to Andorra, Argentina, Aruba, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, British Virgin Islands, Bulgaria, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Faroe Islands, Finland, France, French Polynesia, Germany, Great Britain, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Yemen, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan R.O.C., Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, Uruguay, Venezuela, Vietnam, and Zimbabwe.
Affected count
14,328 ZOOM programmers
Manufactured in
4100 Hamline Ave N, Saint Paul, MN, United States
Products
Boston Scientific Model 3120 ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) installed with Model 2892 Programmer Application Software used for the following pacemaker products: ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus. ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0673-2024Don't want to check this manually?