FDA-Device2025-12-03Class II

IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Catalog Number: 490549; Primary UDI-DI: 04046955270323; Unit of Dose UDI-DI: 04046955270316; Expiration Date: Earliest Exp of Comp or 36mths. 2. Catalog Number: 490565; Primary UDI-DI: 04046955275601; Unit of Dose UDI-DI: 04046955275595; Expir

B Braun Medical Inc
Hazard

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Sold states
Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Affected count
65,232 units
Manufactured in
824 12th Ave, N/A, Bethlehem, PA, United States
Products
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Catalog Number: 490549; Primary UDI-DI: 04046955270323; Unit of Dose UDI-DI: 04046955270316; Expiration Date: Earliest Exp of Comp or 36mths. 2. Catalog Number: 490565; Primary UDI-DI: 04046955275601; Unit of Dose UDI-DI: 04046955275595; Expiration Date: Earliest Exp of Comp or 36mths.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0675-2026

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief