FDA-Device2024-01-17Class II
BrightView, Gamma Camera System, Product Code 882480.
Philips North America
Hazard
While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.
Sold states
Domestic distribution nationwide. Foreign distribution worldwide.
Affected count
359 US; 328 OUS
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Products
BrightView, Gamma Camera System, Product Code 882480.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0677-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief