FDA-Device2016-01-27Class II

GE Healthcare, Aestiva MRI, 1006-9310-000. This version of the Datex-Ohmeda 7900 ventilator is used in Datex-Ohmeda Aestiva/5 anesthesia systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The Aestiva/5 MRI anesthesia system provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment.

GE Medical Systems, LLC
Hazard

The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.

Sold states
Worldwide Distribution - US (nationwide including PR except NH, VT, & WY. Internationally to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA, BRAZIL, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, CHANA, GIBRALTAR, GREECE, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LITHUANIA,MALAYSIA, MALDIVES, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORWAY, OMAN, PAKISTAN, PERU, PHILIPPINES, POLAND, QATAR, REUNION, ROMANIA, RUSSIAN FEDERATION SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI lanka, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.
Affected count
84
Manufactured in
3000 N Grandview Blvd, N/A, Waukesha, WI, United States
Products
GE Healthcare, Aestiva MRI, 1006-9310-000. This version of the Datex-Ohmeda 7900 ventilator is used in Datex-Ohmeda Aestiva/5 anesthesia systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The Aestiva/5 MRI anesthesia system provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0679-2016