FDA-Device2019-12-25Class II
A1CNow SELF CHECK (European), Software Version Revision D, REF Numbers 3054, 3062
Polymer Technology Systems
Hazard
A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.
Sold states
The products were distributed to the following US states: AR, CA, FL, IL, IN, KY, MN, NC, NJ, NY, OH, PA, RI TN, UT, VA, WA, and Puerto Rico. The products were distributed to the following foreign countries: Australia, Bahamas, Belgium, Bahrain, Bangladesh, Canada, China, Columbia, Czech Republic, England, Egypt, Greece, Spain, India, Indonesia, Haiti, Hong Kong, Italy, Japan, Jordan, Kenya, Lebanon, Mexico, Myanmar, Malaysia, Philippines, Pakistan, South Africa, Nigeria, Poland, Saudi Arabia, Sweden, Turkey, Singapore, Trinidad & Tobago, Slovenia, Vietnam, United Arab Emirates, Uruguay.
Affected count
500
Manufactured in
4600 Anson Blvd, Whitestown, IN, United States
Products
A1CNow SELF CHECK (European), Software Version Revision D, REF Numbers 3054, 3062