FDA-Device2024-12-25Class II
Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory electrocardiogram.
Spacelabs Healthcare, Ltd.
Hazard
Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
Sold states
U.S.: AR, CA, CO, DE, FL, FA, LA, MA, MI, MN, MT, NC, NE, NU OH, OR, PA, Puerto Rico, TN, TX, UT, VA, WA, and WI. O.U.S.: Argentina Denmark Kuwait Serbia Australia Dominican Republic Lebanon Singapore Austria Ecuador Lithuania Spain Belgium El Salvador Martinique Sweden Benin France Mexico Switzerland Bolivia Germany Netherlands Taiwan Brazil Greece Oman Trinidad and Tobago Brunei Darussalam Guatemala Panama Turkey Canada Honduras Peru United Arab Emirates Chile Hong Kong Poland United Kingdom China Iceland Portugal Columbia India Venezuela Congo Ireland Qatar Viet Nam Costa Rica Israel Republic of Macedonia Croatia Italy Roumanie Czech Republic Korea, Republic of Saudi Arabia
Affected count
13,202 devices
Manufactured in
Unit B, Foxholes Centre; John Tate Road, Hertford, United Kingdom
Products
Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory electrocardiogram.
Official notice