FDA-Device2016-12-07Class III
Procotyl (Rim- Lock) Liners: REF PHA04702, REF PHA04704, REF PHA04706, REF PHA04708, REF PHA04710, REF PHA04712, REF PHA04714 For use with the conserve metal head
MicroPort Orthopedics Inc.
Hazard
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.
Sold states
Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.
Affected count
500 units
Manufactured in
5677 Airline Rd, Arlington, TN, United States
Products
Procotyl (Rim- Lock) Liners: REF PHA04702, REF PHA04704, REF PHA04706, REF PHA04708, REF PHA04710, REF PHA04712, REF PHA04714 For use with the conserve metal head