FDA-Device2016-12-07Class III
Lineage Shell: REF 36693646, REF 36693848, REF 36694050, REF 36694252, REF 36694454, REF 36694656, REF 36694858, REF 36695060, REF 36695262, REF 36695464 For use with the conserve metal head
MicroPort Orthopedics Inc.
Hazard
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.
Sold states
Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.
Affected count
22 units
Manufactured in
5677 Airline Rd, Arlington, TN, United States
Products
Lineage Shell: REF 36693646, REF 36693848, REF 36694050, REF 36694252, REF 36694454, REF 36694656, REF 36694858, REF 36695060, REF 36695262, REF 36695464 For use with the conserve metal head