FDA-Device2016-02-17Class II

Moog Curlin Infusion, Ambulatory Infusion Pump The Curlin Ambulatory Infusion System includes Curlin Ambulatory multi-therapy and PainSmart pumps. Curlin 4000, catalog number 340; Curlin 6000, 6000 CMS, and PainSmart IOD, catalog number 360; and Curlin Pump Calibration Set, catalog number 340-4038.

MOOG Medical Devices Group
Hazard

Moog is recalling Curlin 4000, 6000, and PainSmart Infusion Pumps manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015 because the master pumps used to calibrate them were out of tolerance and may cause an over delivery of medications and solutions at a rate of 1.2 - 6.8% faster than programmed.

Sold states
Worldwide Distribution - US (nationwide), Guam, and countries of: Australia, Canada, Croatia, Turkey, New Zealand, Saudi Arabia, Japan, Korea, and Italy.
Affected count
11,472
Manufactured in
4314 S Zevex Park Ln, N/A, Salt Lake City, UT, United States
Products
Moog Curlin Infusion, Ambulatory Infusion Pump The Curlin Ambulatory Infusion System includes Curlin Ambulatory multi-therapy and PainSmart pumps. Curlin 4000, catalog number 340; Curlin 6000, 6000 CMS, and PainSmart IOD, catalog number 360; and Curlin Pump Calibration Set, catalog number 340-4038.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0688-2016