FDA-Device2019-12-25Class II
Arrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit, Product Code 9018P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
Teleflex Medical
Hazard
The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
Sold states
US nationwide distribution.
Affected count
526,000 total
Manufactured in
3015 Carrington Mill Blvd, Morrisville, NC, United States
Products
Arrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit, Product Code 9018P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0688-2020Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief