FDA-Device2016-02-10Class I

Emergency Transport Ventilator Oxylog 2000 Plus Ventilator Catalog Number: 5705080. Volume controlled emergency and transport ventilator with pressure support.

Draeger Medical, Inc.
Hazard

Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.

Sold states
Nationwide Distribution -- including states of AR, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Affected count
0
Manufactured in
3135 Quarry Rd, N/A, Telford, PA, United States
Products
Emergency Transport Ventilator Oxylog 2000 Plus Ventilator Catalog Number: 5705080. Volume controlled emergency and transport ventilator with pressure support.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0691-2016