FDA-Device2016-02-10Class I

Emergency Transport Ventilator Oxylog 3000 Ventilator and Oxylog 3000 plus Ventilator Catalog Numbers: 2M86300/2M86965 and 5704813/5704831. Volume controlled and pressure-controlled emergency and transport ventilators.

Draeger Medical, Inc.
Hazard

Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.

Sold states
Nationwide Distribution -- including states of AR, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Affected count
402 units (314 - Oxylog 3000; 88 - Oxylog 3000 Plus)
Manufactured in
3135 Quarry Rd, N/A, Telford, PA, United States
Products
Emergency Transport Ventilator Oxylog 3000 Ventilator and Oxylog 3000 plus Ventilator Catalog Numbers: 2M86300/2M86965 and 5704813/5704831. Volume controlled and pressure-controlled emergency and transport ventilators.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0692-2016