FDA-Device2022-03-09Class II
AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850
Beckman Coulter Inc.
Hazard
Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.
Sold states
US: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming; and OUS: Albania, Algeria, American Samoa, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Ethiopia, Finland ,France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Lesotho, Libya, Lithuania, Macau, Macedonia, Malawi, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Niger, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saint Pierre and Miquelon, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, Vietnam, Yemen, Zambia, Zimbabwe.
Affected count
181,776 Kits
Manufactured in
250 S Kraemer Blvd, Brea, CA, United States
Products
AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0692-2022Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief