FDA-Device2025-12-03Class IIPROCESSING DEFECT

IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 3 CARESITE LADS, 131 IN.; Catalog Number: 490274.

OTHERNationwide distributionMedical devices
Hazard

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Affected count
98,568 units
Manufactured in
824 12th Ave, N/A, Bethlehem, PA, United States
Products
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 3 CARESITE LADS, 131 IN.; Catalog Number: 490274.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0698-2026