FDA-Device2019-01-16Class III
IntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or old
Boston Scientific Corporation
Hazard
Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.
Sold states
Japan
Affected count
1,282 (all three types in Japan)
Manufactured in
300 Boston Scientific Way, Marlborough, MA, United States
Products
IntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system
Official notice