FDA-Device2024-12-25Class II

OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.

GE OEC Medical Systems, Inc
Hazard

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Sold states
U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
Affected count
84 systems
Manufactured in
384 N Wright Brothers Dr, Salt Lake City, UT, United States
Products
OEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic Xray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0703-2025