FDA-Device2018-03-07Class II

digital x-ray detector ProGrade R1 - solid state X ray imager (flat panel/digital imager) As a part of a radiographic system, the Philips ProGrade is intended to acquire, process, store, display, and export digital radiographic images. The Philips ProGrade is suitable for all routine radiographic examinations, including specialist area like intensive care, trauma, or pediatric work, excluding mammography. United States only: The Eleva Workspot is not intended for fluoroscopy and angiograph

Philips Electronics North America Corporation
Hazard

It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the SkyPlate detector by removing its batteries, but the acquired image is lost and a re-take is necessary.

Sold states
Worldwide Distribution - USA (nationwide) Distribution, Internationally to Canada.
Affected count
157
Manufactured in
3000 Minuteman Rd, N/A, Andover, MA, United States
Products
digital x-ray detector ProGrade R1 - solid state X ray imager (flat panel/digital imager) As a part of a radiographic system, the Philips ProGrade is intended to acquire, process, store, display, and export digital radiographic images. The Philips ProGrade is suitable for all routine radiographic examinations, including specialist area like intensive care, trauma, or pediatric work, excluding mammography. United States only: The Eleva Workspot is not intended for fluoroscopy and angiography.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0704-2018