FDA-Device2020-12-30Class II
31" (78 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3011; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3020; ChemoLock Port w/Spiros, Red Cap, REF: CL-35; 31" (79 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Purple Cap, Hanger, REF: CL3951; 40" (102 cm) Appx 7.7 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Drip Chamber, 0.2 Micron
Hazard
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Sold states
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Affected count
88700
Manufactured in
951 Calle Amanecer, N/A, San Clemente, CA, United States
Products
31" (78 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3011; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3020; ChemoLock Port w/Spiros, Red Cap, REF: CL-35; 31" (79 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Purple Cap, Hanger, REF: CL3951; 40" (102 cm) Appx 7.7 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Drip Chamber, 0.2 Micron Filter, Spiros w/Red Cap, Bag Hanger, REF: CL3962; 31" (79 cm) Appx 3.5 ml, Admin Set w/ChemoLock Bag Spike with Additive Port, 20 Drop In-Line Chamber w/15 Micron Filter, Spiros, Purple Cap, Hanger, REF: CL3999
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0707-2021More ICU Medical, Inc. recalls
- FDA-Device2026-07-22ICU Medical Plum Duo Infusion Pump recalled for potential overdose risk
- FDA-Device2026-07-22ICU Medical Plum Solo Precision IV Pump v1.0.3 recalled for potential overdose risk
- FDA-Device2026-06-17ICU Medical Plum Duo Precision IV Pump recalled for missing soft limit alerts
- FDA-Device2026-06-17ICU Medical Plum Duo Infusion Pump recalled for missing soft limit alerts
- FDA-Device2026-06-17ICU Medical Plum Solo Precision IV Pump recalled for missed soft limit alerts
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief