FDA-Device2014-01-22Class II
King Systems KLTSD414 Quantity, 5 PER CASE KING LTSD KIT SIZE 4 King LT(S)-DTM supraglottic airway Identification Part Number(s): KLTSD414 KING LTSD KIT/KIT/TROUSSE/KIT SIZE 4 Contents: 1 - KLTSD Size 4......(155 - 180cm) .. 1 - 60cc Syringe 1 - Sterile Lubricant 32-5332 05/10 Intended for use in adult patients (in excess of 25 kg) for controlled ventilation during anesthesia for procedures that are short in duration.
King Systems Corp.
Hazard
KLTD214 airways found inside KLTSD414 packages. The KLTSD414 airway has a lumen for use of a gastric suction tube; the KLTD214 airway does not include a lumen for a gastric suction tube.
Sold states
Distributed in the states of CA, NY
Affected count
108 cases( 540 products)
Manufactured in
15011 Herriman Blvd, Noblesville, IN, United States
Products
King Systems KLTSD414 Quantity, 5 PER CASE KING LTSD KIT SIZE 4 King LT(S)-DTM supraglottic airway Identification Part Number(s): KLTSD414 KING LTSD KIT/KIT/TROUSSE/KIT SIZE 4 Contents: 1 - KLTSD Size 4......(155 - 180cm) .. 1 - 60cc Syringe 1 - Sterile Lubricant 32-5332 05/10 Intended for use in adult patients (in excess of 25 kg) for controlled ventilation during anesthesia for procedures that are short in duration.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0709-2014Don't want to check this manually?