FDA-Device2020-12-30Class II

APEX Compounding System Control Panel Module, REF: 601238, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.

B Braun Medical, Inc.
Hazard

There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the ordered volume for an ingredient. This could result in a a compounded drug that does not meet the specified accuracy range for an individual ingredient.

Sold states
US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, IL, LA, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, PA, TN,TX, WA.
Affected count
16 units
Manufactured in
1601 Wallace Dr Ste 150, N/A, Carrollton, TX, United States
Products
APEX Compounding System Control Panel Module, REF: 601238, Product Code AX1000 - Product Usage: intended for intravascular administration to a patient.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0719-2021