FDA-Device2021-02-03Class II
ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system
PHILIPS HOME HEALTHCARE SOLUTION
Hazard
After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command.
Sold states
Nationwide
Affected count
91
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Products
ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system