FDA-Device2020-01-01Class II
Cordis S.M.A.R.T. CONTROL /S.M.A.R.T. Vascular Stent System. The S.M.A.R.T. CONTROL and S.M.A.R.T. Vascular Stent Systems consist of a self-expanding stent made of Nitinol (nickel-titanium alloy) material that is pre-mounted on an over-the-wire delivery system. The stents contain tantalum markers at each end. The S.M.A.R.T. CONTROL and S.M.A.R.T. Vascular Stents are designed to open to a preprogrammed diameter at body temperature. There are two delivery system configurations S.M.A.R
Cardinal Health Inc.
Hazard
Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".
Sold states
Worldwide distribution - US Nationwide including Puerto Rico and countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaiian, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Repulic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Maldives, Malta, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saint-Barthelemy, Saudia Arabia, Sinapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela.
Affected count
267,247 units total (262,641 OUS, 4,606 US)
Manufactured in
7000 Cardinal Pl, Dublin, OH, United States
Products
Cordis S.M.A.R.T. CONTROL /S.M.A.R.T. Vascular Stent System. The S.M.A.R.T. CONTROL and S.M.A.R.T. Vascular Stent Systems consist of a self-expanding stent made of Nitinol (nickel-titanium alloy) material that is pre-mounted on an over-the-wire delivery system. The stents contain tantalum markers at each end. The S.M.A.R.T. CONTROL and S.M.A.R.T. Vascular Stents are designed to open to a preprogrammed diameter at body temperature. There are two delivery system configurations S.M.A.R.T. CONTROL and S.M.A.R.T. The stent is contained within the outer sheath of the delivery system. Once the distal end of the delivery system reaches the treatment site, the outer sheath of the delivery system is retracted to expose the stent and start its self-expansion. For the S.M.A.R.T. CONTROL Stent System, sheath retraction is achieved by grasping the handle in a fixed position with the tuning dial held between the thumb and index fingers, and rotating the tuning dial in a clockwise direction until the distal end of the stent is visibly apposed to the vessel wall. For the S.M.A.R.T. Stent System, sheath retraction is achieved by grasping the inner shaft in a fixed position and moving the outer sheath proximally relative t the inner shaft.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0730-2020Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief