FDA-Device2022-12-28Class II
B125P Patient Monitor (with E-module slot option), REF 6160000-002
GE Medical Systems, LLC
Hazard
GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that is not at or near sea level.
Sold states
Worldwide distribution.
Affected count
388 units
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Products
B125P Patient Monitor (with E-module slot option), REF 6160000-002