FDA-Device2019-01-23Class II

LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for

Waldemar Link GmbH & Co. KG (Mfg Site)
Hazard

The reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.

Sold states
US Nationwide in the states of: FL, GA, IA, IN, KS, MI, MN, and TN.
Affected count
11
Manufactured in
Oststr. 4-10, Norderstedt, Germany
Products
LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for the following conditions: Revision arthroplasty due to juxta-articular bone defects Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone Revision of loosened femoral prosthesis components with periprosthetic/subprosthetic fracture Deformed proximal femur due to fractures or osteotomies Correction of bone deficiencies, e.g. due to tumors Large post-revision and post-trauma segmental bone defects Oncological and revision surgery from tibial to hip area (in conjunction with Endo-Model¿ SL Rotational and Hinge Knee Prostheses) The device is intended for cemented and cementless use.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0733-2019