FDA-Device2016-02-10Class II
ABX PENTRA Triglycerides CP ref. A11A01640 is an in vitro diagnostic assay for the quantitative determination of triglycerides in human serum and plasma based on an enzymatic colorimetric test. It is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.
Horiba Instruments Inc
Hazard
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
Sold states
U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
Affected count
216
Manufactured in
9755 Research Dr, N/A, Irvine, CA, United States
Products
ABX PENTRA Triglycerides CP ref. A11A01640 is an in vitro diagnostic assay for the quantitative determination of triglycerides in human serum and plasma based on an enzymatic colorimetric test. It is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.