FDA-Device2022-12-28Class II

HeartWare Model 1102

Heartware, Inc.
Hazard

Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing.

Sold states
Worldwide distribution.
Affected count
892 units
Manufactured in
14400 Nw 60th Ave, Miami Lakes, FL, United States
Products
HeartWare Model 1102

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0744-2023

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief