FDA-Device2022-03-23Class II

Diagnostic Ultrasound System, ACUSON Juniper, UDI: (01)04056869152059, Model: 11335791, Rx, CE Ultrasound imaging system intended for the following applications: Vascular, Abdominal, Neonatal, Echocardiographical, Pediatric, Small Parts, Musculoskeletal, Obstetrical, Gyneological, Intraoperative, and Urological applications using different ultrasound transducers for different applications.

Siemens Medical Solutions USA, Inc.
Hazard

The clip store function in the ultrasound imaging system does not work when the system has a disk full error. This could cause a delay in treatment if the ultrasound system is unable to save clips as study documentation during a high risk procedure, such as a stress echo exam.

Sold states
Worldwide distribution: US (nationwide) and countries of: Algeria, Andorra, Argentina, Australia, Austria, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Faroe Islands, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kosovo, Kuwait, Latvia, Liechtenstein, Lithuania, Luxembourg, Malaysia, Mexico, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Norway, Oman, Peru, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay, Vietnam, and Zimbabwe.
Affected count
2846
Manufactured in
22010 SE 51st St, N/A, Issaquah, WA, United States
Products
Diagnostic Ultrasound System, ACUSON Juniper, UDI: (01)04056869152059, Model: 11335791, Rx, CE Ultrasound imaging system intended for the following applications: Vascular, Abdominal, Neonatal, Echocardiographical, Pediatric, Small Parts, Musculoskeletal, Obstetrical, Gyneological, Intraoperative, and Urological applications using different ultrasound transducers for different applications.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0748-2022