FDA-Device2024-01-24Class II
SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS
Hazard
Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.
Sold states
Worldwide distribution - US Nationwide and the countries of Canada, Europe, Hong Kong, Australia, Japan, Singapore, and Latin America.
Affected count
82,200 total units (5 per box)
Manufactured in
3500 Corporate Pkwy, N/A, Center Valley, PA, United States
Products
SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0748-2024More Olympus Corporation of the Americas recalls
- FDA-Device2026-04-08Olympus SOLTIVE Premium laser system recalled for defective power supply
- FDA-Device2026-04-08Olympus SOLTIVE Pro SuperPulsed Laser System recalled for power supply defect
- FDA-Device2026-04-01Olympus PKS Cutting Forceps recalled for defective welds causing jaw breakage
- FDA-Device2026-04-01Olympus MAJ-1443 and MAJ-1444 endoscope suction valves no longer compatible with OER-Pro and OER-Elite reprocessors
- FDA-Device2026-04-01Olympus PKS Cutting Forceps recalled for defective welds causing jaw breakage
Don't want to check this manually?